Can excess oil clog holding pipettes ?


Yes. Oil droplets can be aspirated if the pipette tip dips too deeply or if oil overlays are uneven: Prevention: Keep the tip just below the aqueous-media interface. Remedy: Back-flush with buffered medium to clear oil before retrying.


Co-Planar Focus: Bring both tips into the same focal plane, then lock Z-axis before adjusting X/Y alignment. Back-Light Contrast: Use oblique illumination to create silhouette outlines—overlap the tips precisely. Test Grasp: Gently approach a microbead in the drop and ensure both pipettes meet at the bead’s center, then back off slightly to confirm symmetrical contact.


Stable Suction: Use just enough negative pressure to hold the zona without deforming the oocyte. Symmetrical Grip: Center the pipette on the widest zona curve, avoiding off-center holds. Oil Overlay: A thin layer of equilibrated oil reduces fluid currents that can spin the embryo.


Check for Leaks: Inspect tubing and o-rings before each cycle. Replace any loose connectors. Pressure Reservoir: Use a syringe or secondary reservoir inline to buffer against rapid pressure drops. Regular Re-priming: Every 10–15 minutes, re-test suction on an oil drop to confirm consistency.


Secure Mounting: Tighten the pipette clamp gently, using vibration-dampening O-rings if available. Thermal Equilibration: Allow the stage and pipette to reach 37 °C stability before alignment to avoid expansion/contraction drift. Periodic Re-check: Every few minutes, briefly touch the tip to the oil drop edge to confirm position.


Forward-Flush Method: Apply brief positive pressure into a medium drop to push bubbles out toward the tip. Back-Flush Option: Reverse the flow gently against a sterile oil interface to coalesce and expel air. Visual Confirmation: Under low magnification, verify a continuous liquid column with no refractive “gaps.”


Typical Range: 15–25° relative to the dish surface, providing a shallow grip without excessive oocyte compression. Adjustment: Fine-tune by lowering in 1–2° increments until the embryo edge seals gently against the tip.


Tilted Zona: The equatorial plane isn’t parallel to the pipette axis—pronuclei won’t face the injection pipette correctly. Off-Center Grip: Embryo wobbles or rotates when suction is applied. Unequal Blastomere Exposure: In cleavage-stage, one blastomere obscures the injection site.


Co-planar Focus: Bring both tips into focus at the same Z-height, then lock that plane. Back-light Contrast: Use oblique illumination to silhouette the tips and overlay them precisely in X/Y. Test Grip: On a microbead or oil droplet, gently close suction to ensure both tips meet centrally before switching to oocytes.


Monash Biotech's Holding Micropipettes are available in two inner diameter sizes: 15µm and 20µm.


Monash Biotech's Holding Micropipettes are available in two outer diameter sizes: 80µm, 100µm and 120µm.


Over-Suction: Excessive negative pressure can disrupt oolemma integrity, causing cytoplasmic leakage. Delayed Handling: Prolonged exposure outside of the incubator increases sensitivity to pressure changes. Zona Fragility: Over-thinned or laser-weakened zona may fail to support even gentle suction, collapsing around the pipette.


The taper length of our Holding Micropipettes is 700 micrometers.


Spike (pointed) Tip: Penetrates thicker zonae for firm grabs Useful in hardened or post-thaw embryos Flat Tip: Larger contact area for uniform suction on fragile blastomeres Reduces local deformation and cell damage


Inner Diameter: 15–20 µm accommodates the larger zona and trophectoderm cells without excessive suction. Outer Diameter: ~25–30 µm for adequate rigidity, minimizing flex during manipulations.


Larger Embryos/Blastocysts: As embryos grow (Day 5+), a longer taper (150–200 µm) offers better reach and stability Thick or Hardened Zona: Longer shank reduces transmitted force at the tip, improving grip on tough zona


Our Holding Micropipettes are available in angles ranging from 15° to 45°, as well as 0° (straight) degrees.


Insufficient Suction: Increase negative pressure in small steps (e.g., +20 hPa). Tip Misalignment: Ensure the pipette mouth sits flush against the zona; angle too steep can lift rather than grip. Zona Hardness Variability: Harder zonae (e.g., post-thaw) may require a slightly wider bore or firmer suction.

FAQ Categories

Click on Your Topic to get answers

Explore More
Intrauterine Insemination
In Vitro Fertilization
Intracytoplasmic Sperm Injection
Injection Micropipettes
Trophectoderm Biopsy (Bevelled) Micropipettes
Trophectoderm Biopsy (Flat) Micropipettes
Zygote Intrafallopian Transfer
History
Gamete Intrafallopian Transfer
Frozen Embryo Transfer
Polar Body Biopsy Micropipettes
Holding Micropipette
Blastomere Biopsy Micropipettes

Got more questions?

Visit our full FAQ Page for in-depth answers or reach out to our Customer Support team.

Monash Biotech Pvt. Ltd.
Change Language
Monash Biotech Pvt. Ltd.

Home

Quality Pledge

About Us

Events Gallery

Blog

News & Updates

Please note that the 3D models displayed on this website are for illustrative purposes only. Actual product dimensions, colors, and finishes may vary.

© 2026 Monash Biotech Pvt. Ltd.. All Rights Reserved.

Designed & Developed by Goafreet Company